Ugrás a tartalomhoz

 

Compensation of the major drawback of oscillometric blood pressure measurement

  • Metaadatok
Tartalom: http://hdl.handle.net/10890/15644
Archívum: Műegyetem Digitális Archívum
Gyűjtemény: 1. Tudományos közlemények, publikációk
Konferenciák gyűjteményei
BME MIT PhD Minisymposium
BME MIT PhD Minisymposium, 2021, 28th
Cím:
Compensation of the major drawback of oscillometric blood pressure measurement
Létrehozó:
Nagy, Péter
Jobbágy, Ákos
Dátum:
2021-07-28T11:37:48Z
2021-07-28T11:37:48Z
2021
Tartalmi leírás:
Oscillometric blood pressure measurement is one of the most commonly used methods to estimate the state of the cardiovascular system. Despite its widespread use, accuracy of the oscillometric method is influenced by several factors and can be questionable in case of arrhythmia, jerky breathing and increased arterial stiffness. This paper analyzes the accuracy of the oscillometric method for measurements of patients affected by cardiovascular diseases. Illustrative examples are given when the conventional oscillometric method produces high errors. A method is proposed for the compensation of these errors based on the photoplethysmographic signal.
Nyelv:
angol
Típus:
könyvfejezet
Formátum:
application/pdf
Azonosító: